USER INPUT FUNCTION
DISPLAY
MODE
DESCRIPTION
USrIN
?
?
NO
Print
P-r5t
Print Request
Print and Reset
Serial transmit of the active parameters
selected in the Print Options menu
(Module 5).
Same as Print Request followed by a
momentary reset of the assigned value(s).
DISPLAY
NO
P-Loc
MODE
No Function
Program Mode Lock-out
DESCRIPTION
User Input disabled.
See Programming Mode Access chart
(Module 3).
rSt-1
rSt-2
rSt12
Setpoint 1 Reset
Setpoint 2 Reset
Setpoint 1 and 2 Reset
Resets setpoint 1 output.
Resets setpoint 2 output.
Reset both setpoint 1 and 2 outputs.
ZErO
Zero Input
(Edge triggered)
Zero the Input Display value causing
Display Reading to be Offset.
? dSP
rESEt
d-HLd
d-SEL
d-LEV
COLOr
Reset (Edge triggered)
Display Hold
Display Select
(Edge Triggered)
Display Intensity Level
(Edge Triggered)
Backlight Color
(Edge Triggered)
Resets the assigned value(s) to the
current input value.
Holds the assigned display, but all other
meter functions continue as long as
activated (maintained action).
Advance once for each activation.
Increase intensity one level for each
activation (backlight version only).
Change backlight color with each
activation (backlight version only).
USER INPUT ASSIGNMENT
U-ASN ? HI          HI-LO
LO dSP
Select the value(s) to which the User Input Function is assigned. The User
Input Assignment only applies if a selection of reset, display hold, or print and
reset is selected in the User Input Function menu.
6.2 MODULE 2 - S ECONDARY F UNCTION P ARAMETERS ( 2-SEC )
2-SEC
SEL
PARAMETER MENU
Pro
HI-En
Max Display
Enable
HI-t
Max Capture
Delay Time
LO-En
Min Display
Enable
LO-t
Min Capture
Delay TIme
FCS
Factory
Service
CodE
Access Code
For Service
Operations
Operations
HI-En ?
?
MAX DISPLAY ENABLE
NO YES
? NO
Enables the Maximum Display Capture capability.
CodE
?
RESTORE FACTORY DEFAULT SETTINGS
Entering Code 66 will overwrite all user settings with
the factory settings. The meter will display rESEt and then
66 return to CodE 00 . Press the SEL button to exit the module.
VIEW VERSION DISPLAY
HI-t
?
?
2.0
MAX CAPTURE DELAY TIME
0.0 to 999.9 seconds
CodE
?
?
50
Entering Code 50 will display the version (x.x) of the
meter. The display then returns to CodE 00 . Press the SEL
button to exit the module.
LO-En ?
?
CodE
? 48
When the Input Display is above the present MAX value for the entered
delay time, the meter will capture that display value as the new MAX reading.
A delay time helps to avoid false captures of sudden short spikes.
MIN DISPLAY ENABLE
NO YES
? NO
Enables the Minimum Display Capture capability.
CALIBRATION
The CUB5P uses stored calibration values to provide
accurate voltage and current measurements. Over time,
the electrical characteristics of the components inside the
meter could slowly change, therefore the stored calibration
values may no longer accurately define the input circuit.
For most applications, recalibration every 1 to 2 years should be sufficient.
Calibration of the CUB5P involves a voltage or current calibration, which
should only be performed by individuals experienced in calibrating electronic
equipment. Allow a 30 minute warm up for eqiupment and unit before
performing any calibration related procedures. The following procedures should
be performed at an ambient temperature of 15 to 35°C (59 to 95°F).
LO-t
?
?
2.0
MIN CAPTURE DELAY TIME
0.0 to 999.9 seconds
CAUTION: The accuracy of the calibration equipment will directly affect the
accuracy of the CUB5P.
Calibration
1. Connect the negative lead of a precision DC source with an accuracy of 0.01%
?
When the Input Display is below the present MIN value for the entered delay
time, the meter will capture that display value as the new MIN reading. A delay
time helps to avoid false captures of sudden short spikes.
FACTORY SERVICE OPERATIONS
FCS ?
NO yES
NO
Select yES to perform either of the Factory Service Operations shown below.
8
or better to the COMM. Leave the positive lead of the DC source unconnected.
2. With the display at CodE 48 , press and hold the SEL button for 2 seconds. Unit
will display CAL NO .
3. Press the RST button to select the range to be calibrated.
4. Press the SEL button. Display reads 0.0A ( 0.0 V for voltage).
5. Apply 0 signal:
Current: leave the positive lead of the DC source unconnected.
Voltage: apply a short to the input or connect the positive lead of the DC
source to INP+ and set the source to 0.
Press SEL . Display reads CALC for about 8 seconds.
6. When the display reads the selected range (10 V, 20 mA, or 50 mA), connect
the positive lead of the DC source to INP+ and apply the full scale input signal
for the range. Press SEL . Display reads CALC for about 8 seconds.
7. Repeat steps 3 through 6 for each input range to be calibrated. When display
reads CAL NO , press the SEL button to exit calibration.
相关PDF资料
PDF描述
CUB5RTR0 RTD METER REFLECTIVE
CUB5TCB0 METER PANEL LCD THERMOCOUPLE RED
CUB5USB0 USB OPTION CARD
CUB5VR00 METER LCD DC VOLT 8DIGIT
CUR-3285-MB BOX ABS 8.25X5.15X3.13" BLACK
相关代理商/技术参数
参数描述
CUB5PR00 功能描述:PROCESS METER REFLECTIVE RoHS:是 类别:工业控制,仪表 >> 仪表 - 面板,数字 系列:* 标准包装:12 系列:* 其它名称:Q7072030
CUB5R000 功能描述:COUNTER/RATE IND DUAL REFL DISPL RoHS:是 类别:工业控制,仪表 >> 计数器 系列:CUB5 其它有关文件:Declaration of Conformity 标准包装:1 系列:99766 计数速率:25Hz 数字/字母数:5 输入类型:机电式脉冲 输出类型:- 电源电压:24V 显示器类型:十进制拨轮
CUB5RLY0 功能描述:BOARD RELAY PLUG-IN FOR CUB5 RoHS:是 类别:工业控制,仪表 >> 配件 系列:* 标准包装:1 系列:Aero-Motive® 130117 附件类型:拖车带 适用于相关产品:标准盒式跟踪系统 其它名称:WM6183
CUB5RTB0 制造商:Red Lion Controls 功能描述:PANEL METER; RTD METER WITH RED/GREEN BACKLIGHT DISPLAY 制造商:Red Lion Controls 功能描述:RTD METER WITH RED/GREEN 制造商:Red Lion Controls 功能描述:TEMPERATURE DISPLAY FOR RTD 制造商:Red Lion Controls 功能描述:PANEL DISPLAY, TEMPERATURE, 95mA, 9V TO 28V; No. of Digits / Alpha:5; Meter Function:Temperature; Meter Range:-19999 to +19999; Digit Height:12.2mm; Panel Cutout Height:39.1mm; Panel Cutout Width:74.9mm
CUB5RTR0 功能描述:RTD METER REFLECTIVE RoHS:是 类别:工业控制,仪表 >> 仪表 - 面板,数字 系列:* 标准包装:12 系列:* 其它名称:Q7072030