参数资料
型号: CUB4TC00
厂商: Red Lion Controls
文件页数: 5/8页
文件大小: 0K
描述: METER PANEL LCD THERMOCOUPLE
标准包装: 1
系列: CUB4
类型: 热电偶
显示器类型: LCD
显示器正面尺寸: 2.95" L x 1.54" W(74.9 x 39.1mm)
显示器位数: 5
显示器位数 - 高度: 0.480"(12.19mm)
背光:
安装类型: 面板安装
端子: 螺钉端子
电源电压: 9 V ~ 26 V
其它名称: Q1421239
Q763911
Voltage Calibration Diagram
1. Connect a precision Voltage Source with an output range of 0.000 to 105.000
mV and an accuracy of 0.03% or better to the SIG+ (#3) and SIG- (#5)
terminals. Placement of jumper does not affect voltage calibration.
2. Enter the Program Menu and step through the Program Menu list by pressing
the “SEL” button until CAL is displayed.
3. Press the “blank” button to change no to YES . Press “SEL”.
4. The display shows 00000. Use the “blank” button to increment the access
code to 00006. Press and hold “SEL”.
5. After entering the correct access code, the unit will display IN1 . Follow the
table below for calibration sequence.The unit displays CALC for about 3
seconds after pressing the “SEL” button before proceeding to the next input
point.
Resistance Calibration Diagram
10. With the display showing r 0 , set the resistance to 0 ohms, then press the
“SEL” button. The unit displays CALC for a few seconds and then the display
will change to r 400 .
11. With the display showing r 400 , set the resistance to 400 ohms, then press
the “SEL” button. The unit displays CALC for a few seconds and then the
display will change to r 10 .
12. Change the input selector jumper to position ‘A’.
13. With the display showing r 10 , set the resistance to 10 ohms, then press the
“SEL” button. The unit displays CALC for a few seconds and then the display
will change to r 0 .
14. With the display showing r 0 , set the resistance to 0 ohms, then press the
“SEL” button. The unit displays CALC for a few seconds and then the display
mV DISPLAY
IN1
IN2
IN3
IN4
IN5
IN6
IN7
IN8
PARAMETER
0.000 mV
15.000 mV
30.000 mV
45.000 mV
60.000 mV
75.000 mV
90.000 mV
105.000 mV
ACTION
Apply 0.000 mV, wait 5 sec., press SEL
Apply 15.000 mV, wait 5 sec., press SEL
Apply 30.000 mV, wait 5 sec., press SEL
Apply 45.000 mV, wait 5 sec., press SEL
Apply 60.000 mV, wait 5 sec., press SEL
Apply 75.000 mV, wait 5 sec., press SEL
Apply 90.000 mV, wait 5 sec., press SEL
Apply 105.000 mV, wait 5 sec., press SEL
will change to r 20 .
15. With the display showing r 20 , set the resistance to 20 ohms, then press the
“SEL” button. The unit displays CALC for a few seconds and then the unit
will return to the CAL program item.
Before exiting the programming menu, verify that the tyPE is set to the desired
type and the input selector jumper is in the appropriate position. Exit
programming and check the calibration. Repeat if necessary
Entering code 85 will place the CUB4RT in a resistance display mode. This
mode is for diagnostic purposes only. When power is removed from the
6. After voltage calibration is complete, the display will show r 200 .
7. Place the input selector jumper in position ‘B’.
8. Connect a precision resistance decade box with an accuracy of ± 0.01 ohms
to the unit as a three wire RTD. See Resistance Calibration Diagram.
9. With the display showing r 200 , set the resistance to 200 ohms, then press
the “SEL” button. The unit displays CALC for a few seconds and then the
display will change to r 0 .
CALIBRATING THE CUB4TC
The CUB4TC uses stored voltage calibration and cold junction temperature
values to provide accurate temperature and voltage measurements. Over time,
the electrical characteristics of the components inside the CUB4TC will slowly
change with the result that the stored calibration values no longer accurately
define the input circuit. For most applications, recalibration every 1 to 2 years
should be sufficient.
Calibration of the CUB4TC involves a voltage calibration and a cold junction
calibration. It is recommended that both calibrations be performed. The voltage
calibration MUST precede the cold junction calibration. Allow 30 minute warm-
up before performing any calibration related procedure. The following
procedures should be performed at an ambient temperature of 15 to 35 °C (59 to
95 °F).
Calibration should only be performed by individuals experienced in
calibrating electronic equipment.
CAUTION: The accuracy of the calibration equipment will directly affect the
accuracy of the CUB4TC.
Voltage Calibration Check
1. Enter the programming menu and set tYPE to LIN and OFSEt to 0 . Exit
programming.
2. Use a precision DC voltage supply with an accuracy of 0.03% or better.
Connect the negative lead of the power supply to TC- (#5) and the positive
lead of the power supply to TC+ (#3).
5
CUB4RT, the display will return to its previously programmed state. If the
RTD type is set for Cu427 and the jumper is set to position A, the display will
read 00.000 to 20.000 ohms. All other RTD types with the jumper in position
B, will cause the display to read 000.00 to 400.00 ohms
3. Compare the CUB4TC read-out to the precision DC supply at various points
over the range (–10.00 mVDC to 60.00 mVDC). The tolerance is ± 2 LSD
(±0.02 mV), at all points within this range.
4. Calibrate the CUB4TC if the readings are out of tolerance.
Cold Junction Calibration Check
1. Enter Programming Menu, and verify the following:
tyPE = probe being connected to the unit.
CJC = YES
dISP = 0.0 ° C
OFSEt = 0
Make any necessary changes to programming.
2. Connect a thermocouple probe of known accuracy and type (J, K, E, T, or N)
to the CUB4TC. The probe should match the one selected in programming.
3. Connect a reference temperature probe to measuring end of thermocouple to
monitor temperature. Allow sufficient time for temperatures to equalize (at
least 15 minutes).
4. Compare unit display with reference temperature indicator. The unit display
should equal the reference probe temperature. Tolerance is ±1.0 °C.
5. Calibrate the cold junction temperature if out of tolerance.
相关PDF资料
PDF描述
CUB5COM1 RS485 SERIAL COMMO CARD
CUB5IB00 DC METER RED/GREEN BKLT
CUB5PB00 PROCESS METER RED/GREEN BKLT
CUB5RTR0 RTD METER REFLECTIVE
CUB5TCB0 METER PANEL LCD THERMOCOUPLE RED
相关代理商/技术参数
参数描述
CUB4V000 功能描述:METER LCD DC VOLT 3.5DIG RoHS:是 类别:工业控制,仪表 >> 仪表 - 面板,数字 系列:CUB4 标准包装:12 系列:* 其它名称:Q7072030
CUB4V-000 制造商:Red Lion Controls 功能描述:Voltage Meter 制造商:Red Lion Controls 功能描述:VOLTAGE METER; No. of Digits / Alpha:3-1/2; Meter Function:DC Millivolts / DC Volts; Meter Range:-199.9mV to +199.9mV / -1.999V to +1.999V / -19.99V to +19.99V / -199.9V to +199.9V; Digit Height:15.2mm; Panel Cutout Height:33.02mm ;RoHS Compliant: Yes
CUB4V010 功能描述:METER LCD DC VOLT 3.5DIG Y/G BCK RoHS:是 类别:工业控制,仪表 >> 仪表 - 面板,数字 系列:CUB4 标准包装:12 系列:* 其它名称:Q7072030
CUB4V020 功能描述:METER LCD DC VOLT 3.5DIG RED BCK RoHS:是 类别:工业控制,仪表 >> 仪表 - 面板,数字 系列:CUB4 标准包装:12 系列:* 其它名称:Q7072030
CUB-51-70010 功能描述:时间延迟和计时继电器 120VAC 1 - 10SEC 10AMP RoHS:否 制造商:Crydom 显示器类型:Hand Dial 电源电压:280 VAC 定时范围: 触点形式: 触点额定值:6 A 端接类型:DIN Rail