参数资料
型号: AP033
厂商: Teledyne LeCroy
文件页数: 21/62页
文件大小: 0K
描述: PROBE DIFFERENTIAL 500MHZ
产品培训模块: Fundamentals of Probing Overview
标准包装: 1
系列: AP
类型: 有源差分
衰减值: 1:1,10:1
带宽: 500MHz
电阻 - 输入: 2M
电压 - 最大: CAT I 42.4V
电容 - 输入: 2.7pF,1.2pF
线缆长度: 42.000"(1066.80mm)
颜色:
包括: AC 耦合器,附件工具包,衰减器,外壳,手册
配用: MSO 44MXS-B-ND - DGTL O-SCOPE MIX SGL 400MHZ DSO
MSO 64MXS-B-ND - DGTL O-SCOPE MIX SGL 600MHZ DSO
MSO 104MXS-B-ND - DGTL O-SCOPE MIX SGL 1GHZ DSO
WAVESURFER 104MXS-B-ND - DGTL O-SCOPE 1GHZ 5GS/S 4CH
WAVESURFER 64MXS-B-ND - DGTL O-SCOPE 600MHZ 5GS/S 4CH
WAVESURFER 62MXS-B-ND - DGTL O-SCOPE 600MHZ 5GS/S 2CH
WAVESURFER 44MXS-B-ND - DGTL O-SCOPE 400MHZ 5GS/S 4CH
WAVESURFER 42MXS-B-ND - DGTL O-SCOPE 400MHZ 5GS/S 2CH
WAVESURFER 24MXS-B-ND - DGTL O-SCOPE 200MHZ 2.5GS/S 4CH
相关产品: PK033-ND - AP033/AP034 ACCES KIT
Operator’s Manual
Figure 4, Effect of Differential Offset
When the AP033 is used with a Teledyne LeCroy oscilloscope equipped with a ProBus interface and
software versions 7.6.0 to 8.0.0, the probe offset must be controlled through either the Probe
Offset menu selection in the COUPLING screen, or the buttons on the probe body. The channel
O FFSET control does not change the probe offset and, therefore, should not be used.
When used with the ADPPS Power Supply on instruments that lack ProBus support, offset can be
controlled with the buttons on the front panel. The offset can be returned to zero at any time by
briefly pressing the ↑ and ↓ O FFSET buttons at the same time. Pressing and holding the ↑ and
↓ O FFSET buttons for more than two seconds will initiate an Autobalance cycle.
Autobalance
The AP033 incorporates an Autobalance function to remove the DC offset from the output.
Autobalance must be invoked by you. When power is first applied, the probe will return to the
internal values resulting from the last Autobalance cycle. For Autobalance to function properly, all
signals must be removed from the input.
After several minutes of warm up, or when the probe is exposed to a large shift in ambient
temperature, some DC offset drift may occur from thermal effects in the amplifier circuitry. To
initiate an Autobalance cycle, remove the probe from the test circuit and select AUTOBALANCE
fr om the oscilloscope’s COUPLING menu for the channel the probe is connected to. If the probe is
being used with the ADPPS power supply, remove the input signal, and push and hold both the
OFFSET buttons for two seconds. The AP033 output must be terminated into 50 Ω for successful
Autobalance.
922260-00 Rev A
15
相关PDF资料
PDF描述
AP034 PROBE ACT DIFFERENTIAL 1GHZ
AP2S2LC OSCILLATOR PROG 1.8V +-50PPM SMD
AP3S1EJ OSCILLATOR PROG 2.5V +-20PPM SMD
APGDT001 SYSTEM DEV LIN SERIAL ANALYZER
ARE0810KS04 SWITCH DIP SPST EXTENDED 8POS SM
相关代理商/技术参数
参数描述
AP033 制造商:Lecroy Corporation 功能描述:DIFF. PROBE 500MHZ ((NS))
AP-033/A/N 制造商:MISCELLANEOUS 功能描述:
AP0332 制造商:SUPERTEX 制造商全称:SUPERTEX 功能描述:8 P-Channel Latchable Power MOSFET Array
AP0332CG 制造商:SUPERTEX 制造商全称:SUPERTEX 功能描述:8 P-Channel Latchable Power MOSFET Array
AP0332ND 制造商:SUPERTEX 制造商全称:SUPERTEX 功能描述:8 P-Channel Latchable Power MOSFET Array