参数资料
型号: AP033
厂商: Teledyne LeCroy
文件页数: 25/62页
文件大小: 0K
描述: PROBE DIFFERENTIAL 500MHZ
产品培训模块: Fundamentals of Probing Overview
标准包装: 1
系列: AP
类型: 有源差分
衰减值: 1:1,10:1
带宽: 500MHz
电阻 - 输入: 2M
电压 - 最大: CAT I 42.4V
电容 - 输入: 2.7pF,1.2pF
线缆长度: 42.000"(1066.80mm)
颜色:
包括: AC 耦合器,附件工具包,衰减器,外壳,手册
配用: MSO 44MXS-B-ND - DGTL O-SCOPE MIX SGL 400MHZ DSO
MSO 64MXS-B-ND - DGTL O-SCOPE MIX SGL 600MHZ DSO
MSO 104MXS-B-ND - DGTL O-SCOPE MIX SGL 1GHZ DSO
WAVESURFER 104MXS-B-ND - DGTL O-SCOPE 1GHZ 5GS/S 4CH
WAVESURFER 64MXS-B-ND - DGTL O-SCOPE 600MHZ 5GS/S 4CH
WAVESURFER 62MXS-B-ND - DGTL O-SCOPE 600MHZ 5GS/S 2CH
WAVESURFER 44MXS-B-ND - DGTL O-SCOPE 400MHZ 5GS/S 4CH
WAVESURFER 42MXS-B-ND - DGTL O-SCOPE 400MHZ 5GS/S 2CH
WAVESURFER 24MXS-B-ND - DGTL O-SCOPE 200MHZ 2.5GS/S 4CH
相关产品: PK033-ND - AP033/AP034 ACCES KIT
Operator’s Manual
B. Incorrect Frequency Response
Possible causes are a defective probe or oscilloscope, poor connections, or poor grounding.
1. Verify that the BW limiting of the oscilloscope is off.
2. Connect the probe to another oscilloscope. If the probe now measures properly, the problem
may be in the oscilloscope.
3. If the probe behaves as if it is ac-coupled at high frequency, check for an open input
connection.
4. Poor frequency or transient response and AC gain errors may result when one of the two input
connections is open.
5. Excessive “ring” and other transient problems can result from excessive input lead length. To
test this, shorten the input leads to less than 1 cm. If the transient response changes
significantly, the lead parasitics are the cause.
C. DC Errors
Incorrect DC gain requires recalibration or factory repair. This can be determined by completing the
gain checks in the Performance Verification Procedure.
1. Extremely high source resistance will result in DC gain errors. Check the probe accuracy with
the oscilloscope calibrator signal.
2. Verify that the probe is not being overdriven into clipping for its current gain setting.
3. Excessive offset can result from large changes in ambient temperature. Remove the input
signal from the probe and repeat the Autobalance cycle. With the Offset set to zero, did the
trace return to the center of the graticule?
D. Poor Common Mode Rejection
Use the 1-kHz calibrator signal from the oscilloscope to check common mode rejection. With both
the + input and – input connected to the calibrator signal, a flat line at zero volts should be seen on
the graticule.
1. Check the probe with the plug-on attenuator installed and removed. If excessive common
mode signal appears only when the attenuator is present, the attenuator may need to be
rematched to the probe. Use the procedure listed in this section to match the attenuator.
2. If the common mode signal appears when the probe is connected to the test circuit, but not
when it is attached to the calibrator, the problem may be caused by large mismatches in the
source impedance. Try connecting both inputs to one of the input signals in the test circuit,
then the other. If the common mode signal disappears, try probing lower impedance points
within the circuit.
922260-00 Rev A
19
相关PDF资料
PDF描述
AP034 PROBE ACT DIFFERENTIAL 1GHZ
AP2S2LC OSCILLATOR PROG 1.8V +-50PPM SMD
AP3S1EJ OSCILLATOR PROG 2.5V +-20PPM SMD
APGDT001 SYSTEM DEV LIN SERIAL ANALYZER
ARE0810KS04 SWITCH DIP SPST EXTENDED 8POS SM
相关代理商/技术参数
参数描述
AP033 制造商:Lecroy Corporation 功能描述:DIFF. PROBE 500MHZ ((NS))
AP-033/A/N 制造商:MISCELLANEOUS 功能描述:
AP0332 制造商:SUPERTEX 制造商全称:SUPERTEX 功能描述:8 P-Channel Latchable Power MOSFET Array
AP0332CG 制造商:SUPERTEX 制造商全称:SUPERTEX 功能描述:8 P-Channel Latchable Power MOSFET Array
AP0332ND 制造商:SUPERTEX 制造商全称:SUPERTEX 功能描述:8 P-Channel Latchable Power MOSFET Array